Fault
|
Possible cuases
|
Solutions
|
The scaling tip doesn’t
vibrate when stepping
on the foot switch
|
The plug in loose or wrong contact
|
Connect Again
|
There is something wrong with
detachable Foot Pedal
|
Connect Again
|
Inside the host
insurance pipe break
|
Inside the host
insurance pipe break
|
Scaler tip doesn’t vibrate but has water comes out
|
Scaler tip loosed
|
Confirm connection of
Scaler tip(See Picture 5)
|
There is fault connection between wire and electronic circuit
|
Contact our company or dealer
|
Handpiece have problem
|
Pull out the Handpiece and send it back to us
|
Wire Have problem
|
Contact our company or dealer
|
The scaling tip vibrates
but there is no water
flowing out
|
Water supply of the
dental unit is off
|
Check the water supply of the dental unit and water adjustable button
|
Electromagnetic
valve with impurities
|
Contact our company or dealer
|
Water plugging
|
Use three water Syringe to dredge the water plug
|
There is still water comes
out after turn off the Unit
|
Electromagnetic
valve with impurities
|
Contact our company or dealer
|
Handpiece become hot
|
Water adjustable
button adjusted too
slow and small the water
|
Adjust the water adjustable button more bigger
|
The water flow is too
small
|
water pressure is too small
|
Increase the water pressure
|
The water pipe of cable is jammed
|
Use three way Syringe to
dredge the water plug
|
The vibration of the ti
becomes weak
|
The tip hasn't been
screwed tightly
|
Screw the scaling tip tightly(See Picture 5)
|
The tip is too loose
which caused by vibration
|
Screw the scaling tip tightly(See Picture 5)
|
There is some water
between the handpiece and the
connector of cable*
|
Dry the connct point
|
The tip is damaged
|
Change it
|
There is water seeping
from the coupling between the handpiece and cable
|
The waterproof O ring
is damaged*
|
Change it
|
No display on screen
|
LCD screen surface
has bruising and other
phenomenas.
|
Suspended for use and
contact our company or dealer.
|
|
Check the main unit a
loose anomaly or display screen connected insecurely.
|
Suspended for use and
contact our company or dealer.
|
|
The power switch was
Damaged
|
Cannot use and contact our company or dealer.
|
No warm water output
|
The red wavy line below the thermometer flicker-free display
|
Please refer to 2.4 operating methods,If still no respond, please contact our company or dealer.
|
|
“Heating” button
become invalid.
|
Please contact our
company or dealer.
|